16.11.2014 Views

Accusetter Manual - BC MacDonald & Co.

Accusetter Manual - BC MacDonald & Co.

Accusetter Manual - BC MacDonald & Co.

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Auto Air Setup<br />

NOTE: The air check(s) mag and zero must be properly setup and the gage<br />

calibrated before setting up the auto air shutoff.<br />

1) From Run mode press the ENTER button and then rotate the ENTER button until<br />

"SYSTEM" is displayed and then press ENTER.<br />

2) Rotate the ENTER button until "AUTO AIR" is displayed and then press ENTER.<br />

3) Rotate the ENTER button until "5", "10", or "15" seconds is displayed and press<br />

ENTER to select the time delay before auto air shutoff.<br />

4) Press "RUN" to return to run mode.<br />

5) After the programmed delay the air will shutoff and the display on the column will<br />

flash "WAIT" for several seconds and then display "AIR OFF". NOTE: During the<br />

time that the display is flashing "WAIT" DO NOT reinsert the air tool into the<br />

part.<br />

6) Close the flow control on the air shutoff valve so that a minimal amount of air is<br />

allowed to flow to the air tooling.<br />

7) Place the air tooling in the least material condition master. For an air plug the LMC<br />

master is the max master and for an air ring the LMC master is the min master.<br />

8) If the column readout detects two light bars of deviation when the air tool is loaded in<br />

the master it will automatically turn on the full air flow to the gage and a<br />

measurement can be made. If the column readout does not detect two light bars of<br />

deviation when the air tool is loaded in the master then slowly open the flow control<br />

valve on the air shutoff valve until a two bar deviation is detected and the air turns on.<br />

9) Remove the air tool from the LMC master. After the programmed delay the air will<br />

shutoff and the display on the column will flash "WAIT" for several seconds and then<br />

display "AIR OFF". NOTE: During the time that the display is flashing<br />

"WAIT" DO NOT reinsert the air tool into the part.<br />

10) Load the air tool back in the LMC to ensure the air turns on. If it does not turn on<br />

open the flow control until the air turns on.<br />

11) Repeat steps 9 and 10 until the air turns on and off correctly. NOTE: If the gage is<br />

used to measure parts with greater clearance than the LMC master it may be<br />

necessary to open the flow control valve more to ensure that the column detects a two<br />

light bar deviation when the air tool is loaded.<br />

4-26

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!