10.07.2015 Views

PXR3 Instruction Manual - Coulton Instrumentation

PXR3 Instruction Manual - Coulton Instrumentation

PXR3 Instruction Manual - Coulton Instrumentation

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

PV offset (Setting range: -10 to 10%FS)[Description]• With this function, predetermined value is added to the inputreading. This parameter is used for adjusting <strong>PXR3</strong>’sindication so that it becomes same as the one of the recorder.• The <strong>PXR3</strong> operates at the displayed PV (the value towhich the PV offset value is added).[Setting example] Adding the PV offset value of 5°C to the input value of 1200 °CDisplayPV indicationSV indicationPV indication1.2.3.4.5.6.Press and hold the SEL key for three seconds.will be displayed.Press the key to display .Press the SEL key once.The current setting ( ) will be displayed.Operating procedurePress the or keys to flicker and to display .Press the SEL key once. will be displayed and will be registered for , and then offset5˚C will be added to the input value. After that, the controller will operate so that the value towhich the offset value of 5˚C is added can be brought close to the set value.(Repeat the procedure from 3 to 5 to check the set value.)If you want to display the operation status, press and hold the SEL key for two seconds.The SV will be displayed on the display area.If unoperated state continues, the PV will be displayed.39

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!