Define primary<br />

analysis settings<br />

<strong>Applied</strong> <strong>Biosystems</strong> SOLiD 4 <strong>System</strong> <strong>SETS</strong> <strong>Software</strong> <strong>User</strong> <strong>Guide</strong><br />

Chapter 2 Prepare Run Settings<br />

Set analysis parameters 2<br />

The Secondary analysis (sequence data alignments to a reference<br />

sequence) will not occur without a preloaded reference sequence.<br />

Perform secondary and tertiary analysis with the BioScope<br />

<strong>Software</strong>. See “Load a Reference Sequence” on page 37.<br />

In <strong>SETS</strong>, you can define the traffic light thresholds as a part of your<br />

setup for primary analysis.<br />

To modify the threshold settings:<br />

1. Log into <strong>SETS</strong>.<br />

2. Go to Main Menu<br />

�Analysis�Primary<br />

Analysis Settings.<br />

3. Click [New...] to begin<br />

modifying the threshold settings.<br />

Note: Click to delete an existing setting<br />

or to edit one.<br />

4. Enter a new value for the Low or High percentage setting for<br />

any of the following parameters:<br />

– Stringency<br />

– Panel alignment<br />

– HQ bead<br />

– Bead retention<br />

23