Load a Reference Sequence<br />

Specify a New<br />

Reference<br />

<strong>Applied</strong> <strong>Biosystems</strong> SOLiD 4 <strong>System</strong> <strong>SETS</strong> <strong>Software</strong> <strong>User</strong> <strong>Guide</strong><br />

Chapter 2 Prepare Run Settings<br />

Load a Reference Sequence 2<br />

If a specific reference sequence is not in the Reference Genome<br />

drop-down list (Secondary Analysis Settings view), you can load a<br />

new reference sequence before you start a run. Open the <strong>SETS</strong><br />

software and perform steps 1 to 4 below to add a new reference<br />

sequence name to the list of choices you see in <strong>SETS</strong>.<br />

Secondary analysis (sequence data alignments to a reference<br />

sequence) would not occur without a preloaded reference sequence.<br />

IMPORTANT! For more complex genomes (such as the human<br />

genome), barcoded samples, and larger reference (and reads) the<br />

secondary analysis should be performed off-instrument with the<br />

BioScope <strong>Software</strong>. Refer to BioScope <strong>Software</strong> for Scientists<br />

<strong>Guide</strong>: Data Analysis Methods and Interpretation (PN 4448431).<br />

If a sample contains more than the maximum allowable number of<br />

reads, the <strong>SETS</strong> software selects only up to the maximum allowed<br />

number of reads and the rest will not be included in the results of the<br />

secondary analysis.<br />

IMPORTANT! Do not install BioScope <strong>Software</strong> on the SOLiD 4<br />

Analyzer.<br />

The Secondary analysis (sequence data alignments to a reference<br />

sequence) will not occur without a preloaded reference sequence.<br />

Perform secondary and tertiary analysis with the BioScope<br />

<strong>Software</strong>. See “Load a Reference Sequence” on page 37.<br />