Applied Biosystems SOLiD™ 4 System SETS Software User Guide ...
Applied Biosystems SOLiD™ 4 System SETS Software User Guide ...
Applied Biosystems SOLiD™ 4 System SETS Software User Guide ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Auto-Correlation<br />
Report<br />
<strong>Applied</strong> <strong>Biosystems</strong> SOLiD 4 <strong>System</strong> <strong>SETS</strong> <strong>Software</strong> <strong>User</strong> <strong>Guide</strong><br />
Chapter 4 View Reports<br />
View Analysis Reports 4<br />
With quality values, it is possible to predict the expected number of<br />
mismatches to the reference genome caused by measurement errors<br />
as depicted in the bar chart above. The blue line shows the<br />
accumulative read proportion with predicted mismatches. For this<br />
run example, about 28% of the color space calls are predicted to have<br />
0 to 0.5 mismatches. About 80% of the bases are predicted to have<br />
approximately 0 to 3 mismatches.<br />
The final graph included with the Quality Values Report shows the<br />
proportions of six groups of quality values versus the primers and<br />
cycles in the run. The proportion of the highest quality bases tend to<br />
decrease in later cycles and increase after the primer resets. (PAC1<br />
represents Primer A Cycle 1, PBC5 represents Primer B Cycle 5).<br />
Use this report to view the correlation between pairs of color calls in<br />
the raw reads.<br />
Click Auto Correlation Report (Run Details�Analysis Reports).<br />