Error Profiles<br />

Report<br />

<strong>Applied</strong> <strong>Biosystems</strong> SOLiD 4 <strong>System</strong> <strong>SETS</strong> <strong>Software</strong> <strong>User</strong> <strong>Guide</strong><br />

Chapter 4 View Reports<br />

View Analysis Reports 4<br />

The Same data as the one on the previous page but sorted in the order<br />

of the chemistry cycles on the sequencing instrument.<br />

Light blue shading or dark red shading shows stronger correlations<br />

between the two cycles. Such correlations indicate potential<br />

problems with swapped reagents, incomplete cleavage, and other<br />

sequencing chemistry problems.<br />

The Error Profiles Report gives the frequency of errors as a function<br />

of base position in the read and type of error produced for a<br />

particular tag.<br />

Click the Error Profiles Report link (Run Details Analysis Reports)<br />

to view the graphs that show the number of mismatches to the<br />

reference sequence, shown by position on read (top) and by cycle<br />

(bottom). Mouse over to view the calculated values for each bar.<br />