Chapter 4 View Reports<br />

View overall run reports<br />

Bead definitions<br />

Bead Type Definition<br />

Usable Beads Beads that, in a particular ligation cycle, are mapped in<br />

each of the four images. Typically, most beads in a<br />

panel are usable.<br />

The ones that are not lie near the edge of the panel,<br />

which for current purposes is defined as the area<br />

imaged by the corresponding reference or focal map<br />

image.<br />

The ligation fluorescence images must be shifted a few<br />

to tens of pixels to register them with the reference<br />

image. If, for example, a ligation image must be shifted<br />

10 pixels to the left to register with the reference image,<br />

beads within 10 pixels of the right edge of the panel will<br />

not be mapped in that ligation image.<br />

Another common case is a panel-cycle where one of<br />

the ligation images fails to register. In this case, none of<br />

the beads are usable.<br />

Good Beads Usable beads that satisfy all of the following conditions:<br />

intensity >0.05<br />

0.0125 < angle ≤0.05<br />

Best Beads Usable beads that satisfy all of the following conditions:<br />

intensity >0.05<br />

angle ≤0.0125<br />

68 <strong>Applied</strong> <strong>Biosystems</strong> SOLiD 4 <strong>System</strong> <strong>SETS</strong> <strong>Software</strong> <strong>User</strong> <strong>Guide</strong>