Chapter 4 View Reports<br />
View overall run reports<br />
Bead definitions<br />
Bead Type Definition<br />
Usable Beads Beads that, in a particular ligation cycle, are mapped in<br />
each of the four images. Typically, most beads in a<br />
panel are usable.<br />
The ones that are not lie near the edge of the panel,<br />
which for current purposes is defined as the area<br />
imaged by the corresponding reference or focal map<br />
image.<br />
The ligation fluorescence images must be shifted a few<br />
to tens of pixels to register them with the reference<br />
image. If, for example, a ligation image must be shifted<br />
10 pixels to the left to register with the reference image,<br />
beads within 10 pixels of the right edge of the panel will<br />
not be mapped in that ligation image.<br />
Another common case is a panel-cycle where one of<br />
the ligation images fails to register. In this case, none of<br />
the beads are usable.<br />
Good Beads Usable beads that satisfy all of the following conditions:<br />
intensity >0.05<br />
0.0125 < angle ≤0.05<br />
Best Beads Usable beads that satisfy all of the following conditions:<br />
intensity >0.05<br />
angle ≤0.0125<br />
68 <strong>Applied</strong> <strong>Biosystems</strong> SOLiD 4 <strong>System</strong> <strong>SETS</strong> <strong>Software</strong> <strong>User</strong> <strong>Guide</strong>