Filter Fasta fails<br />
<strong>Applied</strong> <strong>Biosystems</strong> SOLiD 4 <strong>System</strong> <strong>SETS</strong> <strong>Software</strong> <strong>User</strong> <strong>Guide</strong><br />
Chapter 3 Monitor the Run in <strong>SETS</strong><br />
Troubleshoot failed jobs 3<br />
The Filter_Fasta job details will show a red icon indicating that the<br />
job has failed.<br />
This problem can occur either at the Filter Fasta stage or at the<br />
Barcording stage due to a bad hard-drive, a lack of disk space, or<br />
other system-related issues.<br />
The software displays an error message and provides the Bead ID or<br />
Sequence number where the record is corrupted.<br />
Example<br />
Based on the original system-related issue, you will see an error<br />
message indicating that the “Fasta is empty” or the “Format is<br />
not right expecting sequence. Current line - >1_10_117_F3.”<br />
Click OK to continue.<br />
The error message will contain the problematic file’s location and<br />
record information.<br />
To resolve, you need to fix any system-related issues first. Then, any<br />
of the following step may be applicable:<br />
Check the corrupted Fasta File manually; if possible fix it from<br />
the command line.<br />
Reanalyze primary analysis stage through <strong>SETS</strong> or manually by<br />
changing the database record/flag.<br />
If the SPCH files are also corrupted, report the error message to<br />
the <strong>Applied</strong> <strong>Biosystems</strong> Field Support Engineer.<br />
Note: Report any critical error messages to the<br />
<strong>Applied</strong> <strong>Biosystems</strong> FSE.<br />