Determine the optimal titration point<br />

<strong>Applied</strong> <strong>Biosystems</strong> SOLiD 4 <strong>System</strong> <strong>SETS</strong> <strong>Software</strong> <strong>User</strong> <strong>Guide</strong><br />

Chapter 4 View Reports<br />

View overall run reports 4<br />

Determine which titration has the highest titration metric. This<br />

titration is the optimal titration point.<br />

Determine the bead deposition density for a sequencing run<br />

1. Calculate the concentration of P2-positive beads using the<br />

formula below, where X is the volume of templated beads used<br />

for the WFA sample (equivalent to 15 million beads) and P2# is<br />

given by the WFA report:<br />

2. Use the calculated concentration in place of the value<br />

determined by the NanoDrop ® ND-1000 Spectrophotometer for<br />

more accurate deposition densities when preparing slides for<br />

sequencing. It is recommended to use this resulting calculated<br />

bead concentration to determine the volume of beads for<br />

deposition and multiply that volume by 120%:<br />