CETAC M-7600 Mercury Analyzer Operator's Manual
CETAC M-7600 Mercury Analyzer Operator's Manual
CETAC M-7600 Mercury Analyzer Operator's Manual
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
5 Maintaining the<br />
<strong>Mercury</strong> <strong>Analyzer</strong><br />
Routine maintenance of the analyzer consists of daily and weekly cleaning of<br />
specific analyzer components. Routine maintenance also includes checking for<br />
leaks or other damage. Additional periodic maintenance tasks may be<br />
required, including replacement of the following analyzer components:<br />
peristaltic pump tubing, sample probe, rinse station tubing, GLS capillaries,<br />
GLS drain tubing and Perma Pure ® dryer cartridge.<br />
CAUTION<br />
Discharge static buildup and ground to the analyzer base or cabinet before<br />
performing any maintenance. Avoid touching the contacts on the communication<br />
ports.<br />
Maintenance Schedule<br />
Daily Maintenance (Always Check Before Analysis)<br />
‣ Ensure the autosampler rinse bottle is rinsed between analytical batches<br />
with small amounts 10% HNO 3 follow by DI water and refilled with<br />
acidified rinse solution. For concentration of standards and samples<br />
greater than 20 µg/L, a 5% HCl / 2% HNO 3 v/v should be sufficient. For<br />
trace to ultra-trace calibrations (such as 0 to 500 ng/L) a rinse solution of<br />
1% HCl / 1% HNO 3 v/v should be sufficient.<br />
‣ Ensure the rinse bottle tubes are completely submerged in rinse solution.<br />
The rinse station supply tubing should be at the bottom of the rinse bottle<br />
and the rinse station return tubing should be at the top of the rinse bottle.<br />
This will ensure that the rinse is a true recirculating rinse. Inspect the<br />
rinse station flow and ensure that the rinse is not removed via the sample<br />
probe faster than it is supplied. If the rinse station is being drained faster<br />
than the supply rate, you may need to adjust the autosampler pump rate in<br />
the method or change rinse pump tubing. Replace the autosampler rinse<br />
pump tubing periodically for best performance (See the Autosampler<br />
Operator’s <strong>Manual</strong>).<br />
103