26.11.2014 Views

CETAC M-7600 Mercury Analyzer Operator's Manual

CETAC M-7600 Mercury Analyzer Operator's Manual

CETAC M-7600 Mercury Analyzer Operator's Manual

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

5 Maintaining the<br />

<strong>Mercury</strong> <strong>Analyzer</strong><br />

Routine maintenance of the analyzer consists of daily and weekly cleaning of<br />

specific analyzer components. Routine maintenance also includes checking for<br />

leaks or other damage. Additional periodic maintenance tasks may be<br />

required, including replacement of the following analyzer components:<br />

peristaltic pump tubing, sample probe, rinse station tubing, GLS capillaries,<br />

GLS drain tubing and Perma Pure ® dryer cartridge.<br />

CAUTION<br />

Discharge static buildup and ground to the analyzer base or cabinet before<br />

performing any maintenance. Avoid touching the contacts on the communication<br />

ports.<br />

Maintenance Schedule<br />

Daily Maintenance (Always Check Before Analysis)<br />

‣ Ensure the autosampler rinse bottle is rinsed between analytical batches<br />

with small amounts 10% HNO 3 follow by DI water and refilled with<br />

acidified rinse solution. For concentration of standards and samples<br />

greater than 20 µg/L, a 5% HCl / 2% HNO 3 v/v should be sufficient. For<br />

trace to ultra-trace calibrations (such as 0 to 500 ng/L) a rinse solution of<br />

1% HCl / 1% HNO 3 v/v should be sufficient.<br />

‣ Ensure the rinse bottle tubes are completely submerged in rinse solution.<br />

The rinse station supply tubing should be at the bottom of the rinse bottle<br />

and the rinse station return tubing should be at the top of the rinse bottle.<br />

This will ensure that the rinse is a true recirculating rinse. Inspect the<br />

rinse station flow and ensure that the rinse is not removed via the sample<br />

probe faster than it is supplied. If the rinse station is being drained faster<br />

than the supply rate, you may need to adjust the autosampler pump rate in<br />

the method or change rinse pump tubing. Replace the autosampler rinse<br />

pump tubing periodically for best performance (See the Autosampler<br />

Operator’s <strong>Manual</strong>).<br />

103

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!