Chapter 3: Installing the <strong>Analyzer</strong><br />
To Test the Autosampler<br />
1 In the QuickTrace software, click the Instrument Control button.<br />
Figure 3-70<br />
Instrument Control Button.<br />
2 Click the Autosampler button.<br />
Figure 3-71<br />
Autosampler Button.<br />
3 Click the buttons to turn the autosampler's peristaltic rinse pump on and off,<br />
move the probe (“sipper”) up and down, and move the probe to a sample vial<br />
and back.<br />
Refer to the QuickTrace software online help for more information on how to<br />
control the autosampler.<br />
Step 12: Fill the Rinse Solution Bottle<br />
1 Fill the rinse bottle with trace metal grade 1% HCl / 2% HNO 3 v/v.<br />
NOTE:<br />
When analyzing samples and standards of high concentration such as 20 ppb<br />
or greater use a stronger concentration of acid; such as 5% HCl / 2% HNO 3 or a<br />
similar mix but do not exceed 10% HNO 3 .<br />
2 Insert the rinse solution drain tube (from the autosampler rinse station drain<br />
port, via the peristaltic pump) about ¼ of the height of the rinse bottle.<br />
3 Insert the rinse solution supply tube (to the autosampler rinse station intake<br />
port, via the peristaltic pump) so that it is just above the bottom of the rinse<br />
bottle.<br />
The end of the drain tube must remain above the end of the supply tube.<br />
For help identifying the rinse solution drain and supply tubes, see Figure 3-10<br />
on page 30.<br />