CETAC M-7600 Mercury Analyzer Operator's Manual
CETAC M-7600 Mercury Analyzer Operator's Manual
CETAC M-7600 Mercury Analyzer Operator's Manual
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Operator’s <strong>Manual</strong><br />
Chapter 4: Using the <strong>Analyzer</strong><br />
6 Verify that the sample capillary (inlet insert) is 0.5mm above the Gas-Liquid<br />
Separator (GLS) center post.<br />
7 Open the vents on the waste container.<br />
8 Inspect the peristaltic pump tubing for wear and flat spots (replace if<br />
necessary). If you remove any of the peristaltic pump tubes, do so one channel<br />
at a time, and be careful to return the tube to the appropriate pump channel.<br />
Do not lock the pressure shoes at this time.<br />
9 Place the reagent capillary in a beaker of DI water.<br />
10 Lock down the peristaltic pump pressure shoes.<br />
11 Inspect liquid flows. The GLS drain should be flowing smoothly with no build up<br />
or pulsing of liquid. The waste tube from the peristaltic pump to the waste<br />
container should be liquid/gas etc… with no vibration. If this is not the case<br />
upon inspection, stop immediately and change GLS drain tube and or waste tube.<br />
12 Wet the GLS center post as described on page 84.<br />
13 Inspect the rinse station for a convex liquid bubble adhering to the sample<br />
probe. If this is not the case, change the rinse pump peristaltic tubing.<br />
14 Open the appropriate worksheet (see the online help in the QuickTrace<br />
software) and set the gas pressure to match the method.<br />
You can also set the gas pressure by clicking on the Set Gas & Pump Speed icon.<br />
Figure 4-15<br />
Setting the Gas Pressure.<br />
15 Zero the QuickTrace M-<strong>7600</strong> using the auto zero.<br />
Figure 4-16<br />
Auto-Zero.<br />
90