<strong>CETAC</strong> QuickTrace M-<strong>7600</strong> <strong>Mercury</strong> <strong>Analyzer</strong><br />
Chapter 3: Installing the <strong>Analyzer</strong><br />
2 mm<br />
Figure 3-23<br />
Probe Position with Z-Drive in Highest Position.<br />
13 Center the probe over the rinse station and rotate the rotor counterclockwise<br />
to lower the probe into the rinse station.<br />
14 Use the provided spiral wrap to secure the sample tube to the Z-drive cable<br />
about 5 cm (2 inches) above the top of the Z-drive. There should be just a little<br />
curve to the free sample tube when the probe is lowered, and a loop when the<br />
probe is raised.<br />
The sample tube should naturally curve away from the probe so that it won't<br />
rub or get caught. If necessary, loosen the probe and rotate it so that the<br />
sample tube looks like the pictures below:<br />
Spiral Wrap<br />
Slight Curve<br />
Away from<br />
Probe<br />
Loop<br />
Figure 3-24<br />
Securing the Sample Tube.<br />
37