<strong>CETAC</strong> QuickTrace M-<strong>7600</strong> <strong>Mercury</strong> <strong>Analyzer</strong><br />

Chapter 3: Installing the <strong>Analyzer</strong><br />

Step 10: Connect the Peristaltic Pump on the<br />

<strong>Mercury</strong> <strong>Analyzer</strong><br />

Figure 3-53<br />

Connected.<br />

Peristaltic Pump on the <strong>Mercury</strong> <strong>Analyzer</strong> After Tubing is<br />

Installing the Peristaltic Pump Tubing<br />

1 Release all four peristaltic pump channel clamps.<br />

2 Locate the peristaltic pump tubing in the completion kit. There are two sets of<br />

pump tubing.<br />

‣ The tubes with the yellow stops are for the sample and drain channels.<br />

‣ The tubes with the black stops are for the reagent channel.<br />

57