Operator’s <strong>Manual</strong><br />

Chapter 5: Maintaining the <strong>Mercury</strong> <strong>Analyzer</strong><br />

‣ Inspect the sample peristaltic pump tubing for fatigue and wear. Replace if<br />

too worn or fatigued.<br />

‣ If the pump tubing was left clamped overnight, install new tubing.<br />

‣ Pre-wet the GLS center post and be sure it remains completely wet during<br />

operation.<br />

‣ Check that the liquid flows, to and from the GLS, are smooth. Verify by<br />

close inspection the inlet to the GLS center post and drain exit points.<br />

‣ Be sure the waste bottle will not overflow during the run.<br />

‣ Check that the reagent bottle is sufficiently full for the number of samples<br />

to analyze.<br />

‣ Check that the SnCl 2 is fresh and not precipitated, crystallized, yellowed,<br />

or oxidized or that the small cap on reagent bottle was left open overnight.<br />

Replace if necessary.<br />

‣ For autosampler maintenance, see the appropriate autosampler Operator’s<br />

<strong>Manual</strong>.<br />

Weekly Maintenance<br />

‣ Remove the GLS and clean if residue is building up. See page 112 for<br />

instructions.<br />

‣ Clean the SnCl 2 reagent bottle weekly or before refilling.<br />

‣ Change the pump tubing if it is too worn, appreciably “flattened,” or left in<br />

place overnight.<br />

‣ Empty the waste bottle. Cap all Luer fittings to carry this bottle.<br />

‣ Check the cells and cell windows for cleanliness.<br />

Monthly Maintenance<br />

‣ Clean the GLS. See page 112 for instructions.<br />

‣ Clean the cells and cell windows. See page105 and following.<br />

‣ Replace the GLS inlet tubing and capillary insert. See page 115.<br />

‣ Replace the GLS drain tube. See page 116.<br />

‣ Check that the Perma Pure ® dryer cartridge is still good. A failing dryer<br />

cartridge may be indicated by loss of mercury absorbance sensitivity and<br />

an increase in the baseline of more than 3000 µabs during a short run of<br />

30 minutes or less. If the mercury absorbance for a given standard<br />

solution drops to 50% or more of its original value, change the cartridge.<br />

See page 117.<br />

Yearly Maintenance<br />

‣ Replace the Perma Pure ® dryer cartridge bi-yearly, or as needed. (See page<br />

117.) A failing dryer cartridge may be indicated by loss of mercury<br />

absorbance sensitivity and an increase in the baseline of more than<br />

3000 µabs during a short run of 30 minutes or less. If the mercury<br />

104