<strong>CETAC</strong> QuickTrace M-<strong>7600</strong> <strong>Mercury</strong> <strong>Analyzer</strong><br />
Chapter 4: Using the <strong>Analyzer</strong><br />
RANGE #2: M-<strong>7600</strong> µg/L<br />
0.1 – 20 µg/L Hg<br />
Gas Flow<br />
100 mL/min<br />
Peristaltic Pump Speed 50%<br />
Sample Flow Rate<br />
~5 mL/min<br />
ASX Rinse Pump Speed 50%<br />
Sample Time (for Liquid Uptake or autosampler “Sip”) 40 s<br />
Rinse Time<br />
95 s<br />
Read Delay<br />
52 s<br />
Replicate Read Time<br />
1.5 s<br />
Replicates 4<br />
Baseline Correction Method 1 point (20-25 s)<br />
Expected Results: 10 ppb | 7% HCl<br />
~150,000 µAbs<br />
Detection Limit (nominal):<br />
< 0.01 ppb<br />
Sample Throughput Rate (minutes/sample)<br />
~ 2.25 min/sample<br />
Table 4-2<br />
µg/L Parameters to Optimize<br />
Figure 4-27<br />
Typical Results from µg/L Settings.<br />
99