<strong>CETAC</strong> QuickTrace M-<strong>7600</strong> <strong>Mercury</strong> <strong>Analyzer</strong><br />

Chapter 4: Using the <strong>Analyzer</strong><br />

RANGE #2: M-<strong>7600</strong> µg/L<br />

0.1 – 20 µg/L Hg<br />

Gas Flow<br />

100 mL/min<br />

Peristaltic Pump Speed 50%<br />

Sample Flow Rate<br />

~5 mL/min<br />

ASX Rinse Pump Speed 50%<br />

Sample Time (for Liquid Uptake or autosampler “Sip”) 40 s<br />

Rinse Time<br />

95 s<br />

Read Delay<br />

52 s<br />

Replicate Read Time<br />

1.5 s<br />

Replicates 4<br />

Baseline Correction Method 1 point (20-25 s)<br />

Expected Results: 10 ppb | 7% HCl<br />

~150,000 µAbs<br />

Detection Limit (nominal):<br />

< 0.01 ppb<br />

Sample Throughput Rate (minutes/sample)<br />

~ 2.25 min/sample<br />

Table 4-2<br />

µg/L Parameters to Optimize<br />

Figure 4-27<br />

Typical Results from µg/L Settings.<br />