<strong>CETAC</strong> QuickTrace M-<strong>7600</strong> <strong>Mercury</strong> <strong>Analyzer</strong><br />

Chapter 4: Using the <strong>Analyzer</strong><br />

‣ Click the “Wet The Gas Liquid Separator Post” button. This will result in a<br />

gas flow of about 1000 mL/min with a pump rate of 100%.<br />

Figure 4-6<br />

Wet GLS Button.<br />

3 Check that the bottle supplying the autosampler rinse station is filled with<br />

clean trace metal grade acidified rinse solution.<br />

4 Place the reagent uptake tube in a beaker of DI water.<br />

5 If the peristaltic pump is not already on, turn it on.<br />

Figure 4-7<br />

Turning On the Pump.<br />

6 Engage the pressure shoe (quick release mechanism) on the peristaltic pump.<br />

7 Pinch the drain tube as shown in Figure 4-8. The drain tube runs from the GLS<br />

to the inlet side of the peristaltic pump.<br />

Pinch the Drain tube here<br />

Figure 4-8<br />

Pinching the Drain Tube.<br />