Operator’s <strong>Manual</strong><br />

Chapter 3: Installing the <strong>Analyzer</strong><br />

3 Connect the tee on the left (inlet) side to the drain tube coming from the GLS.<br />

Drain Tube<br />

Figure 3-57<br />

GLS Drain Tube.<br />

From Drain<br />

To Waste<br />

Figure 3-58<br />

Waste and Drain Connections.<br />

4 Find a Luer male barbed fitting (look for this in the 10 liter waste container).<br />

Replace one of the Luer caps on the waste container lid with the barbed fitting.<br />

Attach the other end of the waste tube to this fitting.<br />

To prevent pressure buildup in the 10-liter waste container, be sure that at<br />

least one of the two vents on the 10-liter waste container are open (uncapped)<br />

during operation.<br />