Chapter 5: Maintaining the <strong>Mercury</strong> <strong>Analyzer</strong><br />

Changing the Cell Gas Tubing<br />

Begin by opening the optical access panel as shown on page 106.<br />

Next, replace the tubing. The replacement tubes, which are in the cell gas<br />

tubing kit, are labeled and precut to length. See Figure 5-10 for label<br />

designations. Match these numbers and letters with the existing tubing to see<br />

where each labeled tube should go. It is best to replace one tube at a time.<br />

Figure 5-10<br />

Tubing Diagram and Connection Table.<br />

To remove an old tube from a plastic connector, grip it near its connector and<br />

pull firmly. For glassware, it is better to slit the old tube with a razor blade or<br />

sharp knife, before removing. Alternatively, you may use the edge of a<br />

fingernail to ease the tube off its glass arm.<br />

A simple way to avoid making wrong connections is to remove only one tube at<br />

a time, and replace this tube with the appropriate, labeled new one, before<br />

proceeding to the next tube.<br />

Drying cartridges are replaced as assemblies (replacements with tubing<br />

already attached).<br />

When finished, be sure no tubing is pinched when the covers are replaced.<br />