<strong>CETAC</strong> QuickTrace M-<strong>7600</strong> <strong>Mercury</strong> <strong>Analyzer</strong><br />
Chapter 5: Maintaining the <strong>Mercury</strong> <strong>Analyzer</strong><br />
Changing the Cell Gas Tubing<br />
Begin by opening the optical access panel as shown on page 106.<br />
Next, replace the tubing. The replacement tubes, which are in the cell gas<br />
tubing kit, are labeled and precut to length. See Figure 5-10 for label<br />
designations. Match these numbers and letters with the existing tubing to see<br />
where each labeled tube should go. It is best to replace one tube at a time.<br />
Figure 5-10<br />
Tubing Diagram and Connection Table.<br />
To remove an old tube from a plastic connector, grip it near its connector and<br />
pull firmly. For glassware, it is better to slit the old tube with a razor blade or<br />
sharp knife, before removing. Alternatively, you may use the edge of a<br />
fingernail to ease the tube off its glass arm.<br />
A simple way to avoid making wrong connections is to remove only one tube at<br />
a time, and replace this tube with the appropriate, labeled new one, before<br />
proceeding to the next tube.<br />
Drying cartridges are replaced as assemblies (replacements with tubing<br />
already attached).<br />
When finished, be sure no tubing is pinched when the covers are replaced.<br />
113