<strong>CETAC</strong> QuickTrace M-<strong>7600</strong> <strong>Mercury</strong> <strong>Analyzer</strong><br />
Chapter 3: Installing the <strong>Analyzer</strong><br />
10 Connect the Liquid Mix tube to the small tee which you just installed.<br />
The Liquid Mix tube is a black tube which connects to the GLS, as shown in<br />
Figure 3-50 on page 55.<br />
Liquid Mix<br />
Tube<br />
Figure 3-63<br />
Liquid Mix Connection.<br />
Step 11: Power On and Verify Communication<br />
Once installation of the QuickTrace M-<strong>7600</strong> system is complete, it is<br />
important to verify that the system is installed correctly.<br />
CAUTION<br />
Attempting to use the QuickTrace M-<strong>7600</strong> before ensuring that all components<br />
are installed correctly may result in damage to the system.<br />
To Configure the Network Connection<br />
The <strong>CETAC</strong> M-<strong>7600</strong> mercury analyzer is controlled by a PC. <strong>CETAC</strong><br />
recommends that this PC be equipped with two network cards. One card is<br />
used for communication with your laboratory network, and the other card is<br />
used exclusively for communication with the M-<strong>7600</strong> mercury analyzer.<br />
In most cases, a PC is supplied with the mercury analyzer, and this PC has two<br />
network cards.<br />