Chapter 5: Maintaining the <strong>Mercury</strong> <strong>Analyzer</strong><br />
settled after cleaning. Re-check the surface reflection to see if the window is<br />
completely clean. If residue, fingerprints, or particles remain, repeat the<br />
process with another pre-cleaned alcohol-moistened foam swab until the<br />
windows are clean. If this quick procedure fails, it may be necessary to<br />
dismantle the assembly for more rigorous “total” cleaning, as described below.<br />
Figure 5-4 Cell assembly diagram.<br />
A - Window cap C - Phillips screws E - Cell end cap<br />
B - Window O-ring D - Sapphire windows F - Glass cell<br />