Operator’s <strong>Manual</strong><br />

Chapter 3: Installing the <strong>Analyzer</strong><br />

5 After 30 seconds press the Up button.<br />

6 Measure the water remaining in the cylinder, and calculate the flow rate.<br />

7 Park the sample probe.<br />

NOTE:<br />

When properly adjusted the tension on the bottom three peristaltic pump<br />

tubes (channels 1-3) should be the same and the tension screw for the top<br />

pump tube (reagent tube, channel 1) should be screwed in 1 to 2 mm farther<br />

than the other channels.<br />

Once the clamp tension on the pump tubing is established, relieve their<br />

stretch:<br />

8 Unclamp reagent and sample clamps.<br />

9 Park the sample probe.<br />

10 Press the Up button to remove the sample probe from the rinse station.<br />

11 Remove the reagent uptake tube from the SnCl 2.<br />

12 Allow the waste tube to empty.<br />

13 Unclamp waste tube clamps.<br />

14 Turn the pump off using the software controls.<br />

Do not leave tubes clamped in place when the system is not being used. The<br />

next time the system is used, hook the tubes and close the quick-release<br />

mechanisms. No screw adjustments will be needed. Previous clamp tension is<br />

“remembered” as the quick release is engaged and disengaged.<br />

74