Operator’s <strong>Manual</strong><br />

Chapter 5: Maintaining the <strong>Mercury</strong> <strong>Analyzer</strong><br />

10 Locate the mA ADJ potentiometer and use a small flat-blade screwdriver<br />

(jewelers screwdriver) to adjust the lamp current.<br />

Figure 5-27<br />

Adjusting the Lamp Current.<br />

‣ Begin by rotating the potentiometer fully clockwise until the OVER RANGE<br />

LED lights, then counterclockwise until the lamp current reads < 0.5 mA. At<br />

this point, the lamp is ready for adjustment.<br />

‣ Adjust the lamp current potentiometer until the marker is in the center<br />

(the green region) of the scale.<br />

‣ Clockwise will increase the current and counter-clockwise will decrease<br />

the current.<br />

‣ After each adjustment, click Check Lamp Current.<br />

‣ Let the voltage stabilize and make fine adjustments until the system<br />

stabilizes.<br />

11 Reinstall the cover and begin using the mercury analyzer.<br />

130