CETAC M-7600 Mercury Analyzer Operator's Manual
CETAC M-7600 Mercury Analyzer Operator's Manual
CETAC M-7600 Mercury Analyzer Operator's Manual
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>CETAC</strong> QuickTrace M-<strong>7600</strong> <strong>Mercury</strong> <strong>Analyzer</strong><br />
Chapter 4: Using the <strong>Analyzer</strong><br />
Running the Interactive Demo<br />
The QuickTrace software comes with an interactive demonstration which will<br />
help you quickly understand the basic operation of the software.<br />
1 From the Start menu, select All Programs | <strong>CETAC</strong> QuickTrace | Interactive<br />
Demo.<br />
Figure 4-9<br />
Starting the Interactive Demo.<br />
Figure 4-10<br />
Interactive Demo Interface.<br />
Overview of the <strong>CETAC</strong> QuickTrace Software<br />
The QuickTrace software lets you create a method tailored to your analytical<br />
needs. You can specify a calibration, quality control limits with error actions<br />
and specific end-of-run routines such as automated standby, print or export<br />
routines.<br />
There are three basic parameters that affect the M-<strong>7600</strong>’s method ranges:<br />
‣ Gas flow rate<br />
‣ Sample peristaltic pump rate<br />
‣ Sample uptake rate<br />
By changing these basic parameters, you can set the QuickTrace M-<strong>7600</strong><br />
mercury analyzer to function in method ranges of ultra-trace to high µg/L<br />
87