Mini<strong>RAE</strong> <strong>3000</strong> User’s <strong>Guide</strong>During duty cycling (PID lamp cleaning), the display shows these iconsin alternation:If there is a pump failure or obstruction that disrupts the pump, you willsee this icon blinking on and off:If you see this blinking icon, consult the Troubleshooting section of thisguide.Calibration StatusThe instrument displays this icon if it requires calibration:Calibration is required (and indicated by this icon) if:• The lamp type has been changed (for example, from 10.6 eV to9.8 eV).• The sensor has been replaced.• It has been 30 days or more since the instrument was lastcalibrated.• If you have changed the calibration gas type withoutrecalibrating the instrument.22