Mini<strong>RAE</strong> <strong>3000</strong> User’s <strong>Guide</strong>3. Zero calibration starts a 30-second countdown and displays thismessage:Zeroing...During the zeroing process, the instrument performs the Zerocalibration automatically and does not require any action on your part.Note: To abort the zeroing process at any time and proceed to Spancalibration, press [N/-] at any time while zeroing is being performed.You will see a confirmation message that says “Zero aborted!” and thenthe Span calibration menu appears.When Zero calibration is complete, you see this message:Zeroing is done!Reading = 0.0 ppmThe instrument will then show the Calibration menu on its display, withSpan Calib highlighted.38