13.07.2015 Views

AED 10 User Manual - Welch Allyn

AED 10 User Manual - Welch Allyn

AED 10 User Manual - Welch Allyn

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>User</strong> <strong>Manual</strong> Specifications 75DefibrillatorOutputEnergy Sequence (user configurable)Charge Time from “Shock Advised”Analysis TimeCombined Analysis and Charge Timefrom power upafter 6 shocksAudible PromptsVisual PromptsControlsWaveform DetailsMRL Orbital TM Biphasic Truncated exponentialShock 1: 150 J, 200 JShock 2: 150 J, 200 J, 300 JShock 3: 150 J, 200 J, 300 J, 360 J4 sec. to 150 J8 sec. to 200 J15 sec. to 360 J4-16 sec.From start of analysis to shock ready:Less than 20 seconds to 200 JLess than 30 seconds to 360 J18 audible prompts20 text screen promptsFour buttons - On/Off, Shock, and 2 software-configurable buttonsTable 1 provides details of the biphasic truncated exponential waveformdelivered by the <strong>AED</strong> <strong>10</strong> (set to Emax) when connected to resistive loads of 25through 175 . The waveforms are characterized by typical values for peakcurrent (I p ), duration of the first output phase, and duration of the second outputphase.I p1phase 1phase 2I p2Table 1PatientImpedance ()I p1(Amps)I p2(Amps)t phase1(ms)t phase2(ms)25 52.2 35.7 5.6 3.850 26.5 16.9 11.6 7.775 17.5 11.8 16.6 9.9<strong>10</strong>0 13.3 9.3 18.8 9.9125 <strong>10</strong>.5 7.7 19.9 <strong>10</strong>.3150 8.9 6.8 20.2 <strong>10</strong>.3175 7.5 6.0 20.4 <strong>10</strong>.3

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!