13.07.2015 Views

AED 10 User Manual - Welch Allyn

AED 10 User Manual - Welch Allyn

AED 10 User Manual - Welch Allyn

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

84 Specifications <strong>Welch</strong> <strong>Allyn</strong> <strong>AED</strong> <strong>10</strong> Automated External DefibrillatorResultsepisodes). Three ED50 estimation procedures were run in parallel, with the device beingused alternated on each shock. After each of the three independent ED50 estimationprocedures had been completed, the entire protocol was repeated twice more, each timestarting all devices at an energy of 30 joules. The ED50 peak current and energy was thenestimated for each animal by logistic regression analysis. Individual phase durations andoverall pulse durations were measured and recorded on each shock.The study consisted of 344 total fibrillation/defibrillation episodes. The mean ED50 andID50 estimates (to one decimal place) are shown below. The significance of difference (pvalue)was calculated by the Wald test in each case, and is shown below. Also shown arethe mean total durations measured for each device.Summary table — ED50, ID50 and DurationMeanMonophasicWaveform<strong>Welch</strong> <strong>Allyn</strong><strong>AED</strong> <strong>10</strong>2kV BiphasicWaveformID50 Peak Current (Amps) 9.0 6.4 8.3Significance of difference (p-value)

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!