Front Panel OperationINOUTTriggerLinkLine #12001 MultimeterTrigger INLinkOUT7002 Switch <strong>System</strong> Trigger LinkCable(8501)Figure 4-31Trigger Link connections (semi-synchronous example)The two instruments are configured as follows:<strong>Model</strong> 7002:Scan List = 1!1-1!10,Arm Layer:Arm Spacing = Immediate*Arm Count = 1*Arm Trigger Control = Acceptor*Scan Layer:Scan Spacing = Immediate*Number of Scans = 1Scan Trigger Control = Acceptor*Channel Layer:Channel Spacing = TrigLinkTrigger Link Mode = Semi-SynchronousSemi-Sync Line = #1Trigger Control = Source*Number of Channels = Use Scanlist Length** Indicates that the setting is the RESET (and factory) default condition<strong>Model</strong> 2001:Arm Layer:Arm Source = ImmediateArm Count = 1Arm Trigger Control = AcceptorScan Layer:Scan Source = ImmediateScan Count = 1Scan Trigger Control = AcceptorMeasure Layer:Measure Source = TrigLinkTrigger Link Mode = Semi-SynchronousSemi-Sync Line = #1Trigger Control = AcceptorMeasure Count = 10To run the test, simply press STEP on the <strong>Model</strong> 7002. Thefollowing explanation on operation is referenced to the operationmodel shown in Figure 4-32.4-69