Trigger subsystemIEEE-488 ReferenceDefaultsPower-up*RST:SYSTem:PRESetSaved power-on setup2 (all layers)2 (all layers)Querycommands:OLINe? Query programmed output line #Short-form formats: :arm:tcon:asyn:olin?:arm:lay2:tcon:asyn:olin?:trig:tcon:asyn:olin?Response message: 1 to 6DescriptionThese commands are used to select an output line for the asynchronous Trigger Link of the specifiedlayer. Keep in mind that asynchronous Trigger Link input and output cannot share the sameline. If you assign the output to a line that is already being used for the input, an error will occurand the command will be ignored.Programmingexample10 OUTPUT 707; “:trig:tcon:asyn:olin 4; olin?”20 ENTER 707; A$30 PRINT A$40 ENDLine 10 Assigns the asynchronous Trigger Link output of the Trigger Layer to line #4, andthen queries the programmed output line.Line 20 Addresses the 7002 to talk.Line 30 Displays the programmed output line # (4).:SSYNchronous:LINE :TRIGger[:SEQuence[1]]:TCONfigure:SSYNchronousSpecify semi-synchronous Trigger Link line for the TriggerLayerParameters = 1 Line #1= 2 Line #2= 3 Line #3= 4 Line #4= 5 Line #5= 6 Line #6Short-formformat:trig:tcon:ssyn:line DefaultsPower-up*RST:SYSTem:PRESetSaved power-on setup1 (Trigger Layer)1 (Trigger Layer)5-145