Adjusting Instrument Parameterswhere:DRIFT is the Drift constant in the instrument during readings and (T2-T1) isin units of days.ExampleSelecting: T2 = 15:10:30Gives the corresponding readings:andTherefore,DRIFT = 0.000R2 - R1 = 0.034 mGalT1 = 14:56:27R2 = 3969.418R1 = 3969.382T2 - T1 = 0.23383 hrs =.00974 daysDRIFT' = -3.6949MaintenanceAdjusting the Drift CorrectionThe final step is to apply the calculated driftcorrection. To do this:PRESSF5Select the AUTOGRAV icon in the main SETUPscreen and press the F5 key. The system displays theAUTOGRAV PARAMETER setup screen.PRESSF1Press the F1 key to access thesecond AUTOGRAV PARAMETER setup screen.Press the up and down arrows to navigate to theDrift field.5-15CG-5 Manual - part # 867700 Revision 2