Applying Tilt CorrectionsPRESSF1The system stores the X1 and R1 values for internalcalculation. Press the F1 key to clear the overlay.Next the reading can be either stored or deleted bypressing the RECORD or CANCEL keys. Note thatthe choice here does not affect the internal offsetcalculation.Access the LEVELING screen again:• by pushing the MEASURE key twice if you didnot record the data.• by pushing the MEASURE key once if you didrecord the data.MaintenanceAdjust the tripod until X = -150±10 and Y = 0±5 inthe LEVELING screenPRESSF2Press the F2 key. The systemmeasures until the reading time has elapsed or theF5 key is pressed.PRESSF5Press the F5 key to view the datain the AUTOGRAV FINAL DATA screen. Thesystem then overlays a User Calibration windowwith:• values of tilt and gravity readings• The error in the current offset value• the existing Tilt X. Offset (Old X Offset)• the new calculated Tilt X. Offset (New XOffset).5-23CG-5 Manual - part # 867700 Revision 2