Using the Autograv ScreenDrift:Long term instrument drift due to stress relaxation inthe elastic (quartz) system. Units are in mGal/day.Drift Start TM/DT:Set the drift start time to the current time.Note:The drift start time changes automatically to thecurrent time when the memory is cleared.PRESSF5When finished, press the F5 key. The systemreturns to the previous menu.SetupSaving Changes to ParametersIf you have made changes to either the correctionsand filters or instrument parameters as describedabove, you must save these changes before they areinitiated in your next survey.PRESSF5In the AUTOGRAV PARAMETER SETUP screen,press the F5 key. The changes aresaved.Cancelling Changes to ParametersPRESSF4If you have made changes but decide you want toundo your changes, press the F4 key inthe AUTOGRAV PARAMETER SETUP screen.2-17CG-5 Manual - part # 867700 Revision 2