Maintaining Your CG-5 and Trouble-shootingPress F2 to accept or F3 to reject the New X Offset.If the value is accepted the instrument internalconstants are automatically updated.Note: The difference between the old and the newoffsets should be relatively small in mostcircumstances. Less than say 10 or 20 arcseconds. Ifthe reported error is larger than this there hasprobably been a problem with the measurement.Important:Do not accept the value and repeat the measurement.If the reported error is larger than 100 arcsec andyou have repeated the measurement several timescontact <strong>Scintrex</strong> before accepting the New X Offset.Repeat the measurement of TiltX.Offs to check thatthe error is less than 1 arcsec.Performing TiltY. Offs AdjustmentThis process is analogous to the process forobtaining TiltX. Offs described above.5-24CG-5 Manual - part # 867700 Revision 2