Fit the connections and joints on the traverse device according to the relevant operating manuals.<br />
SQ256 Retraction Device<br />
3 Operation<br />
3.1 General information<br />
not applicable<br />
3.2 Start-up<br />
Warning!<br />
The probe must only be operated by qualified personnel who are familiar with all maintenance<br />
measures. Damage may occur if operated by unauthorized personnel and maintenance is<br />
improperly carried out.<br />
Check accessibility to the retraction device. Make a visual check of the installation of the connecting lines.<br />
Check the fastenings of the runner rail and the probe.<br />
Check the chain tension, chain wheel and rollers and readjust and grease them if necessary. See Maintenance.<br />
Important!<br />
Check that the voltage specified on the nameplate agrees with the mains voltage.<br />
Check the direction of rotation of the electrical and pneumatic motors.<br />
If the three-phase motor turns in the wrong direction, two mains supply lines must be interchanged.<br />
When closing the terminal box, give special attention to proper sealing. Regrease the mating points if necessary.<br />
Fit stop-plugs to the unused cable entry holes on the three-phase motor and the limit switches.<br />
Adjust the limit switches by inserting and withdrawing the probe. When doing this, loosen the mounting bracket by<br />
releasing the screws and then moving it to the required position and retightening. Then move the probe electrically<br />
to the limit positions and check the switching of the limit switches. Adjust the limit switches for height if necessary.<br />
If the compressed air motor is being used, check the pneumatic limit switch by withdrawing the probe using<br />
compressed air and adjust it if necessary.<br />
Adjust the probe cover in the inserted state.<br />
Note:<br />
The flat point on the cover must be underneath.<br />