Fit the connections and joints on the traverse device according to the relevant operating manuals.<br />

SQ256 Retraction Device<br />

3 Operation<br />

3.1 General information<br />

not applicable<br />

3.2 Start-up<br />

Warning!<br />

The probe must only be operated by qualified personnel who are familiar with all maintenance<br />

measures. Damage may occur if operated by unauthorized personnel and maintenance is<br />

improperly carried out.<br />

Check accessibility to the retraction device. Make a visual check of the installation of the connecting lines.<br />

Check the fastenings of the runner rail and the probe.<br />

Check the chain tension, chain wheel and rollers and readjust and grease them if necessary. See Maintenance.<br />

Important!<br />

Check that the voltage specified on the nameplate agrees with the mains voltage.<br />

Check the direction of rotation of the electrical and pneumatic motors.<br />

If the three-phase motor turns in the wrong direction, two mains supply lines must be interchanged.<br />

When closing the terminal box, give special attention to proper sealing. Regrease the mating points if necessary.<br />

Fit stop-plugs to the unused cable entry holes on the three-phase motor and the limit switches.<br />

Adjust the limit switches by inserting and withdrawing the probe. When doing this, loosen the mounting bracket by<br />

releasing the screws and then moving it to the required position and retightening. Then move the probe electrically<br />

to the limit positions and check the switching of the limit switches. Adjust the limit switches for height if necessary.<br />

If the compressed air motor is being used, check the pneumatic limit switch by withdrawing the probe using<br />

compressed air and adjust it if necessary.<br />

Adjust the probe cover in the inserted state.<br />

Note:<br />

The flat point on the cover must be underneath.<br />

Copyright ® SIEMENS Page 101 04/04