SQ250 <strong>FLK</strong> <strong>Gas</strong> <strong>Sampling</strong> <strong>System</strong><br />

4.5 Measures to be taken during faults<br />

If a fault occurs and is acquired, the red fault lamp begins to flash quickly. While ever the fault is still present or has not been<br />

acknowledged (fault lamp flashes or is continuously illuminated), the relevant message text can be read in the control cabinet<br />

with the OP3 Operating Panel.<br />

Messag<br />

e<br />

No.<br />

Display text in<br />

operating panel<br />

Cause Effect Remedy<br />

0 Probe Emergency<br />

Stop<br />

Emergency stop button<br />

pressed<br />

Probe does not withdraw by<br />

electric motor, but instead<br />

pneumatically for a fault or<br />

power failure<br />

Important!: Dangerous condition during faults, e.g.<br />

coolant overtemp. cannot withdraw probe, risk of<br />

probe damage. For power failure, probe withdraws<br />

despite emergency-off; reset emergency-off, reset,<br />

insert probe.<br />

1 Control <strong>System</strong><br />

Fuse/Voltage fault<br />

Fuse failed<br />

Probe withdraws<br />

(el. +pneum.)<br />

Find cause of fuse failure and remedy, replace fuse,<br />

reset, insert probe<br />

2 Control <strong>System</strong> ESB<br />

S7 I4.2<br />

Power failure or fuse for<br />

PLC failed<br />

Probe withdraws<br />

pneumatically<br />

For fuse failure, find the cause and rectify it; after<br />

power is restored reset, insert probe<br />

3 Cooling <strong>System</strong> Flow<br />

< min<br />

Coolant flow too low<br />

(delayed 10s)<br />

Probe withdraws<br />

(electr. + pneum.)<br />

Coolant pump stops<br />

Check the coolant level, top up if required, check for<br />

leakage, manual valves must be open, start pump and<br />

vent, blocked oil filter. Reset, insert probe.<br />

4 Cooling <strong>System</strong><br />

Pump fault<br />

Power failure, flow too<br />

low, motor prot. switch<br />

tripped<br />

Probe withdraws<br />

(electr. + pneum)<br />

Find cause of motor prot. switch trip and rectify.<br />

Reset, start pump, insert probe<br />

5 Cooling <strong>System</strong> Fan<br />

fault<br />

Motor prot. switch<br />

tripped<br />

Probe withdraws<br />

(electr. + pneum.)<br />

Find cause of motor prot. switch trip and rectify.<br />

Reset, start pump, insert probe<br />

6 Cooling <strong>System</strong><br />

Level to low<br />

Coolant level too low<br />

Probe withdraws<br />

(electr. + pneum.)<br />

Find cause of coolant lost (leakage) and rectify; switch<br />

off pump, top up coolant, start pump, vent, reset, insert<br />

probe.<br />

Vent again when coolant temperature > 180 °C.<br />

7 Cooling <strong>System</strong><br />

Temperature to high<br />

Coolant temperature<br />

to high (default setting<br />

220°C)<br />

Probe withdraws<br />

(electr. + pneum.)<br />

Check coolant level and flow, vent, clean cooler, check<br />

GROW controller with 3-way valve, check limit<br />

setting on temp. controller. No obstacle in cooling air<br />

path? Rectify fault, reset, insert probe, vent again for<br />

coolant temp. > 170 °C<br />

8 Probe retraction<br />

Plugs not OK<br />

I 9.6 for plug monitoring<br />

(Valve combination not<br />

"1")<br />

Probe withdraws<br />

(electr. + pneum.)<br />

Check plug on valve combination for correct seating; if<br />

still not OK, check wiring, reset, insert probe<br />

9 <strong>Sampling</strong> <strong>Gas</strong><br />

Probe/Filter clog.<br />

Still low pressure in<br />

meas. gas line even after<br />

multiple purging<br />

(pressure switch)<br />

Probe does not withdraw.<br />

"Purging" lamp flashes at a<br />

fast rate<br />

Switch off filter heating fuse F1, start purging 3x to<br />

cool down filter, remove filter and clean, clean probe<br />

by undoing rear screw, also check line to condensate<br />

vessel and measuring gas lines, remover filter, ensuring<br />

absolute sealing tightness, switch on fuse F1.<br />

Check for leakage, reset Counter 16, insert probe.<br />

10 Probe retraction ESS<br />

Motor<br />

Either output for<br />

Contactor Insert or<br />

Contactor Withdraw is on<br />

and no feedback or<br />

output not on and<br />

feedback<br />

Probe withdraws<br />

(pneumatically)<br />

Check wiring, especially the limit switches which are<br />

directly latched to the corresponding contactor.<br />

Check outputs and inputs on the PLC:<br />

Output:<br />

Contactor Insert Probe<br />

Input: Feedback<br />

Output: Contactor Withdraw Probe<br />

Input: Feedback<br />

Reset, insert probe<br />

Copyright ® SIEMENS Page 28 04/04