SQ253 Valve Combination<br />
4.3.1 Replacing the valves<br />
• Pull off the cable head after loosening the screw.<br />
• Loosen screwed joints on the housing.<br />
• Unscrew the pipes at both ends.<br />
• Before fitting the new valve<br />
Note:<br />
Pay attention to voltage and type of current according to the nameplate.<br />
• Clean the pipe work of contamination (metal swarf, sealing material).<br />
Note: The arrow indicates the direction of flow. Do not use the valve as a lever when screwing it in. Do not block pilot hole<br />
on the valve output with pipe ends, sealing material, etc. Use PTFE tape as sealing material.<br />
• Assembly takes place in the reverse sequence.<br />
4.3.2 Cleaning the valve<br />
• Pull off the cable head after loosening the screw.<br />
• The pilot valve can be removed after loosening the four screws on the solenoid section.<br />
• Take care with the core and spring.<br />
• Loosen four screws on the valve cover.<br />
• Remove the upper part.<br />
• Take out the diagram and, if necessary, clean it and replace damaged parts.<br />
• The valve body can now be cleaned.<br />
• Grease with silicone grease or acid-free Vaseline.<br />
• Assembly takes place in reverse sequence.<br />
Note:<br />
Observe the tightening torque of 1Nm for the cable head.<br />
4.3.3 Faults:<br />
• Check connections. operating pressure and voltage.<br />
• Pilot hole on the valve output covered or contaminated ?<br />
Solenoid does not pull in:<br />
• Short circuit or coil interruption.<br />
• Core or core aperture contaminated.<br />
• With alternating current, jammed core causes coil overheating.<br />
Leaky valve:<br />
• Procedure as under Section 4.3.2<br />
SQ253 Valve Combination<br />
Copyright ® SIEMENS Page 69 04/04