29.08.2015 Views

FLK Gas Sampling System - MPIP - Free

FLK Gas Sampling System - MPIP - Free

FLK Gas Sampling System - MPIP - Free

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

SQ253 Valve Combination<br />

3. Operation<br />

3.1 General information<br />

This equipment unit has been supplied functionally tested. Therefore, no complicated checks of the separately installed<br />

devices are needed.<br />

3.2 Start-up<br />

Warning! The valve combination in conjunction with the withdrawal device must only be put into operation by<br />

qualified personnel, who are familiar with all maintenance measures. Unauthorized operation and improperly carried out<br />

start-up procedures can cause damage.<br />

The start-up is mainly restricted to a check of the wiring, a functional check and a leakage test of the complete measuring<br />

equipment.<br />

3.2.1 Electrical start-up<br />

After checking the wring to the control cabinet, the following functional checks can be carried out. These should be done<br />

before the compressed air is switched on.<br />

• The individual settings on the four-way valve should be selected from the PLC controller using the OP3<br />

Operating Panel.<br />

• Similarly, the function of the compressed air valves -Y2, -Y3, -Y4, and -Y5 should be checked.<br />

• It must also be checked whether the flushing is initiated and started by the local push buttons. The flushing<br />

process must also be started by simulating low pressure (turning back the limit switch contacts to -300 mbar).<br />

• The flushing sequence must be checked according to the sequence in the operating manual for the controller.<br />

3.2.2 Compressed air and leakage test<br />

With the compressed air it must be ensured that the 6...8 bar pressure, the quantity and the quality conform to the<br />

specification.<br />

Caution:<br />

There should be no one situated in the front area of the probe.<br />

The compressed air must be shut off immediately if any leaks occur.<br />

Turn-on should be carried out slowly and special attention must be paid to the joining pieces.<br />

The compressed air part with the valves should be checked for sealing tightness (use leak-seeking spray or soap-water<br />

solution as an aid).<br />

After starting the flushing, the connecting lines to the filter and the probe should be tested for sealing tightness once the leakseeking<br />

agent has been applied.<br />

If a compressed air motor is being used on the withdrawal device, the function of the withdrawal valves must be checked.<br />

• In the unenergized state the solenoid valve (Figure 2, Item 5) must be open and the compressed air motor driven.<br />

• When the pneumatic limit switch on the withdrawal device is operated, the pneumatic valve (Figure 2, Item 4)<br />

closes and shuts off the compressed air feed to the motor.<br />

Note: The solenoid valve must close absolutely tight when energized, because otherwise the traverse carriage will<br />

move slowly in operation.<br />

Copyright ® SIEMENS Page 66 04/04

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!