SQ253 Valve Combination<br />

2. Installation<br />

2.1 General information<br />

The valve combination is constructed such that it can be easily mounted on the withdrawal device. If no withdrawal device is<br />

being used, then fixing on a mounting frame is possible. Documentation for this can be requested from the SIEMENS<br />

company in the relevant country.<br />

2.2 Mounting on the withdrawal device<br />

The valve combination is fastened to the four screw bolts provided on the opposite side of the drive motor on the traversing<br />

carriage.<br />

All internal connecting screws on the measuring gas line must be retightened, including the nuts on the condensate vessel<br />

(Figure 2, Item 11). Once the valve combination has been mounted, the hose and cable connections to the withdrawal device<br />

and the electrically heated filter must be made.<br />

Note:<br />

See general operating manual Chapter 2.5, Mounting the probe with withdrawal device.<br />

The joining hardware needed for this is included in the supplied items (supplementary pack). All connections have been<br />

labeled to eliminate confusion.<br />

2.2.1 Hose connections to the withdrawal mechanism<br />

These consist of:<br />

Connections between valve combination and electrically heated filter<br />

• Metal-braided hose with R 3/4" threaded joint at both ends for compressed air for the probe flushing.<br />

• Metal-braided hose with R 1/2" threaded joint at both ends for compressed air for the filter flushing.<br />

• Depending on the measuring gas connection between filter and condensate vessel (Figure 2, Items 11,20):<br />

PTFE pressure hose with union nuts (M16x1.5) on both ends.<br />

TEFLON pipe for high-corrosion version with TEFLON screwed glands.<br />

Note: Special attention should be paid to ensure that the PTFE pressure hose or the TEFLON pipe continuously<br />

slopes down to the condensate vessel.<br />

Connections for the compressed air motor. ( Figure 2, View A. ):<br />

• Compressed air feed line to the pneumatic limit switch connection M16x1.5<br />

Valve combination "P2" to the pneumatic limit switch "R1/P2"<br />

• Return line from pneumatic limit switch connection M16x1.5<br />

Valve combination "A" to the pneumatic limit switch "A"<br />

• Compressed air feed line to the compressed air motor connection M16x1.5<br />

Copyright ® SIEMENS Page 64 04/04