SQ253 Valve Combination<br />

Leaky valve:<br />

• Procedure as in Section 4.7.2 or replace the valve.<br />

Fig. 7<br />

Electrically operated compressed air valve<br />

4.8 Pneumatically operated compressed air valve ( optional )<br />

This compressed air valve is only fitted when a compressed air motor is being used for the emergency withdrawal on the<br />

withdrawal device.<br />

The valve has, depending on the choice of connections, the function "no throughput when unenergized" (11 blocked, 2 to 33<br />

vented) or the function "throughput when energized (11 to 2, 33 blocked).<br />

The valve switches only for the duration of the signal on terminal 110 or 12.<br />

4.8.1 Replacing the valve<br />

• Loosen the screwed joints on the housing.<br />

• Pull off the hose after opening the hose clamps on both sides.<br />

• Remove the valve.<br />

• Unscrew the screw-in hose nozzles and the reducing unions.<br />

• Check the type of valve before fitting the new one.<br />

• Clean the hose nozzles, use PTFE tape for sealing the thread.<br />

• Screw in the new valve.<br />

• Assembly takes place in the reverse sequence.<br />

Copyright ® SIEMENS Page 73 04/04