SQ254 Heat Exchanger<br />

3. Setting into operation<br />

3.1 Safety specifications<br />

Be aware of the hot coolant. Ensure, that the coolant temperature is low before opening any connections.<br />

3.2 Preparation before first Start-up<br />

Perform the following steps:<br />

1) visual inspection of the device and hose on tightness<br />

2) check the level of the coolant. It must not be less than the “COLD LEVEL”<br />

3) adjust the flow meter to 1,5 m 3 /h minimum flow<br />

4) open the stop valve (14) entirely<br />

5) set the main switch into the “ON” position<br />

6) press S1 to start the circulation pump<br />

7) adjust the coolant flow by means of the stop valve (14) to 3,6m 3 /h<br />

8) Open the bleed valve (16) for short time so that air can escape from the expansion<br />

vessel<br />

9) observe the coolant through the inspection glass to ensure that no bubbles are in the<br />

coolant<br />

10) Insert the probe to the kiln<br />

11) observe the increase of the coolant temperature<br />

12) at a coolant temperature of 180°C the fan should be switched on<br />

13) at 180°C check the coolant level in the expansion vessel<br />

14) in case of too high “COLD LEVEL” the coolant escapes through the check valves<br />

Copyright ® SIEMENS Page 81 04/04