SQ254 Heat Exchanger<br />

3.3.4 Maximum limit value emergency extraction of probe<br />

The set point of excess temperature is adjusted to 220 °C ex works. As soon as this temperature is reached, the<br />

probe moves out of the furnace.<br />

3.4 Standard starting<br />

Check level of liquid “Level cold” through the inspection glass, if necessary, refill. Check by visual inspection the<br />

tightness of the entire equipment. Turn-on main switc. In case level of cooling liquid has dropped below “Level<br />

cold”, the equipment cannot be set into operation. This malfunction is indicated by the signal lamp H7 “Level min”.<br />

After having refilled liquid, the equipment can be set into operation.<br />

4. Setting out of operation<br />

4.1 Shutdown the heat exchanger<br />

The gas sample probe has to be retracted from the kiln. Keep the heat exchanger in operation for a short time in<br />

order to avoid an accumulation of heat. In case of repair work, switch off the main switch.<br />

5. Maintenance<br />

Weekly:<br />

Check all flanges and hoses on tightness.<br />

If required, tighten screws or replace gaskets.<br />

Check axial seal of pump on tightness.<br />

If required, replace the seal.<br />

Check coolant flow and clean the dirt trap, if necessary.<br />

Monthly:<br />

Check contamination of cooling fins.<br />

In case of severe contaminations, use compressed air for cleaning.<br />

Note: Contamination increases the run-time of the cooling fan.<br />