FLK Gas Sampling System - MPIP - Free
FLK Gas Sampling System - MPIP - Free
FLK Gas Sampling System - MPIP - Free
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
SQ251 <strong>FLK</strong> <strong>Gas</strong> <strong>Sampling</strong> Probe<br />
3 Operation<br />
3.1 General information<br />
Danger !<br />
The probe and coolant lines have an operating temperature of up to 220° Cel.<br />
The baked-on deposits on the probe can reach temperatures of over 1000° Cel.<br />
3.2 Start-up<br />
Warning!<br />
The probe must only be operated by qualified personnel who are familiar with all maintenance<br />
measures. Damage may occur if operated by unauthorized personnel and maintenance is<br />
improperly carried out.<br />
See the operating manual for the complete gas analysis equipment with <strong>FLK</strong> Probe for the start-up procedures.<br />
4 Maintenance<br />
The following maintenance work must be carried out by the user to maintain the probe ready for operation:<br />
When using the probe on the furnace inlet, checks must be made in the initial period during each insertion and<br />
retraction. A check should be made of whether the baked-on deposits on the probe are released by compressed<br />
air during insertion. This opportunity should be used to check the probe for mechanical damage (e.g. due to falling<br />
material deposits in the furnace inlet chamber).<br />
The probe and coolant line connections should also be checked for leaking coolant.<br />
Any leaks must be immediately repaired.<br />
If the baked-on deposits are not released on insertion, then the probe must be cleaned manually.<br />
Danger !<br />
The probe and coolant lines have an operating temperature of up to 220° Cel.<br />
The baked-on deposits on the probe can reach temperatures of over 1000° Cel.<br />
With other installation locations (precalcination chamber) a regular check of the probe for damage, baked-on<br />
deposits and leakage on the coolant line connections must be carried out. The period between these checks<br />
should not exceed 14 days.<br />
If the sampling chamber and/or the sampling pipe becomes blocked, the probe must be cleaned manually. Once<br />
the screwed gland on the sampling side has been removed, the sampling pipe can be cleaned with a suitable rod<br />
or strong wire.<br />
If cleaning is no longer possible, the complete sampling pipe must be changed.<br />
If the ball valve becomes leaky, the tap spindle and/or seals must be replaced.<br />
Copyright ® SIEMENS Page 40 04/04