SQ253 Valve Combination<br />
4. Maintenance<br />
4.1 General information<br />
The user should carry out the following maintenance work so that operational readiness is ensured:<br />
• The valve combination must be thoroughly cleaned of dust and dirt every week.<br />
• The compressed air filter should be cleaned every month.<br />
• A leakage test of the complete system (compressed air side and measuring gas side) must be carried out every six<br />
months.<br />
4.2 Compressed air filter<br />
The compressed air filter is used for the effective separation of liquid and solid particles from the compressed air. However, it<br />
is not intended for oil separation.<br />
The air flows in the direction of the arrow through the filter and is made turbulent by the<br />
guide vane A. The particles of liquid are thrown against the internal walls and run<br />
downwards. The separating cap B forms a calm zone which prevents the separated<br />
condensate from being taken up again by the air flow. Then the air flows through the filter<br />
element C. Here all solid contaminants are retained which are larger than the pore size.<br />
When the condensate reaches a certain height, a valve is opened by the float F and the<br />
filter is automatically drained.<br />
With filters with manual drainage, the filter is drained by operating the drain screw E.<br />
Clean the compressed air filter at least once per month, more often if necessary.<br />
To do this, shut off the compressed air before the filter. Open the filter by turning<br />
the metal container to the right and then clean it.<br />
automatische/manuelle Entleerung<br />
automatic / manual drainage<br />
Fig. 4 Compressed air filter<br />
Copyright ® SIEMENS Page 67 04/04