Siemens<br />

SQ252 Electrically Heated Dedusting Filter<br />

4.4 Replacing the thermostat<br />

• Switch off the fuse for the filter heater or withdraw the plug for the filter on the valve combination.<br />

• Start the compressed air flushing a few times, if available, to cool the filter.<br />

• Loosen the electrical connection cable in the terminal box (Item 12) and pull back the cable.<br />

• Remove the thermostat after loosening the screws at the top of the filter.<br />

• Smear the new thermostat with heat conducting paste on the bottom and firmly screw to the filter.<br />

• Insert the cable into the terminal box and connect according to the wiring diagram.<br />

• Carry out a functional test.<br />

4.5 Measures for fault rectification<br />

4.5.1 Filter does not reach operating temperature after about 15min.<br />

• Check the voltages in the terminal box.<br />

• If there is no voltage on the heater, replace the thermostat.<br />

• If voltage is present on the heater, replace the heater or the complete filter housing.<br />

4.5.2 No release of the measuring gas pump<br />

• Check that the filter operating temperature is about 200 ºCel.<br />

• Check the voltages in terminal box.<br />

• If the temperature is still too low after about 15 min., carry out procedure in 4.4.1.<br />

• If the operating temperature is reached, but thermostat not closed for release, replace the thermostat.<br />

4.5.3 Filter becomes blocked too quickly<br />

If the filter becomes blocked soon after cleaning:<br />

• Check whether the release for the measuring gas pump only occurs after reaching the operating temperature.<br />

• Replace auxiliary-contact thermostat. Procedure as in 4.3.<br />

• Is there enough compressed air available for flushing?<br />

• Check the compressed air pressure and check supply line if necessary.<br />

• Is there compressed air available as required (free of oil and dry) ?<br />

• Check the filter tube for blocked pores (black, burnt deposits).<br />

Note: If oil compounds are present in the compressed air, then baked-on deposits which are difficult to<br />

remove occur in the filter tube. The tube can then only be replaced.<br />

Copyright ® SIEMENS Page 55 04/04