SQ256 Retraction Device<br />
4.2 Guide rollers<br />
If the carriage does not run smoothly or a level of play arises on the carriage when, for example, the probe tip is<br />
moved up and down, then the guide rollers must be readjusted.<br />
Legends:<br />
1 Hex. fastening screws<br />
2 Hex. adjusting screw<br />
3 Lock-nut<br />
Fig. 6 Cross-sectional drawing of carriage<br />
Guide roller adjustment:<br />
• The lower fastening screws on the control box must be removed and the upper screws loosened. Then swivel<br />
the control box upwards.<br />
• On each side of the housing panel on the carriage, the five fastening screws ( 1 ) for the intermediate rail must<br />
be loosened at the bottom.<br />
• Then loosen the lock-nut ( 3 ) for the guide roller adjustment.<br />
• Slightly tighten the adjustment screw so that the lower guide roller is pressed against the traverse rail.<br />
• Traverse the carriage manually over the complete range and make sure that it gives good running<br />
characteristics, readjusting if necessary.<br />
• Tighten the lock nuts.<br />
• Tighten all side fastening screws ( 1 ).<br />
• Correctly fasten the control box again.<br />
Copyright ® SIEMENS Page 103 04/04