SQ253 Valve Combination<br />
4.4 Condensate separator<br />
The condensate separator consists of a glass bulb, lower and upper cover plates in stainless steel each with a connection<br />
opening and O-rings between the glass bulb and covers. Since the condensate vessel is blown out with each flushing process,<br />
no maintenance work is required.<br />
When replacing the various parts, it should be ensured during assembly that the four threaded bolts with hexagon nuts are<br />
tightened evenly and that the O-rings are correctly seated.<br />
If any leaks occur:<br />
• First, slightly, but evenly tighten the threaded bolts.<br />
• Check the glass bulb for any damage.<br />
• Check the O-rings.<br />
Copyright ® SIEMENS Page 70 04/04