26.12.2012 Views

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

UV ABSORPTION METER 153<br />

-_-.-<br />

+-<br />

R - V 0.1 Zero<br />

10MR l00mV Int 1<br />

(6.14)<br />

The potentiometers ‘Zero’ and ‘En’ must be arranged in a position such that during the<br />

analysis the output voltage on ‘Lin’ always remains between 0 and 100 mV. An example<br />

of the linearity of the circuit shown in Fig.6.18 is given in Fig.6.20, in comparison with<br />

that of the older circuits. Most of the isotachopherograms considered in this chapter were<br />

obtained with non-linear circuitry, but the data collected for the various operational<br />

systems that are given in Chapters 11-17 were obtained with linear circuitry.<br />

At a resistance of 1 MR, the accuracy of the resistance determination with linear<br />

circuitry is better than 2 kf2, whde at 10 MR the accuracy is better than 30 kR. If the<br />

ambient temperature increases by 10°C, the difference in resistance determination at a<br />

level of 1 Mi2 is less than 1 kR and at 10 Mi2 it is less than 20 kR.<br />

The stepwise trace obtained during an isotachophoretic analysis when the zones pass<br />

the conductimeter can easily be interpreted for qualitative and quantitative information,<br />

because the steps are sufficiently sharp. This is in contrast with thermometric recording.<br />

Two main reasons can be given why a differentiator was constructed. (1) If small<br />

zones are present, stacked between the others, these zones can be detected much easier<br />

and better with an electronic differentiator. (2) If electronic devices are available and<br />

applied for measuring the time interval between two successive peaks, a pulse is needed<br />

for the printer. A possible circuit for such a differentiator is given in Fig.6.21.<br />

The circuit with IC6 is actually the differentiator. Its amplification can be chosen by<br />

the potentiometer ‘Diff. The time constant is 24 msec, which is small enough to differentiate<br />

the signal on the output ‘En’. The circuit IC7 and lCs is a double-phase rectifier.<br />

A potential recorder with a sensitivity of 100 mV can be applied in combination with<br />

ths differentiator. IC9 forms, together with the two resistors on the non-inverting input,<br />

a Schmitt trigger with a very small hysteresis. The 10-pF capacitor ensures that the<br />

input signal is equal to the first differential of the output signal of the rectifier. If this<br />

value is zero, the Schmitt trigger is triggered. This procedure is shown schematically in<br />

Fig.6.22.<br />

The printed values of the time intervals between the fast moving (about 1 m/h) zone<br />

boundaries increase the accuracy of time recording and hence the quantitative data<br />

handling and, moreover, simplifies the latter. The accuracy in time recording is better<br />

than 10 msec, which represents an accuracy of 5 - g-equiv./l, assuming an electric<br />

driving current of 70 /.LA and a concentration of the leading electrolyte of lo-* g-equiv./l.<br />

6.5. UV ABSORPTION METER<br />

6.5.1. Introduction<br />

Because in the steady state of an isotachophoretic separation all components move in<br />

individual zones with identical speeds and only the counter ion is mixed with these<br />

components, specific detectors such as the W absorption meter can offer much additional

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!