26.12.2012 Views

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

382 COUNTER FLOW OF ELECTROLYTE<br />

Amaranth red was used as the sample in the operational system at pH 6 (Table 12.1).<br />

Various terminators with different effective mobilities were applied. The results are shown<br />

graphically in Fig.17.5.<br />

These experiments indicated that the disturbance is greater;the smaller are the differ-<br />

ences in effective mobilities. In Fig.17.2 (loo%*), two examples of these boundary<br />

disturbances are shown: (a) the boundary when MES was applied as the terminator and<br />

(b) the boundary when glutamate was applied as the terminator. Experiments with<br />

acetate as the terminator showed that the amaranth red is flushed back into the<br />

terminating reservoir. From these series of experiments, we can conclude that a ‘slow’<br />

terminator is needed in order to prevent too much sample from being flushed back.<br />

Fig.17.5. Disturbance of the zone boundary as a function of the effective mobility of the following<br />

zone. ah = Difference in effective mobility; L = length of the disturbance of the zone boundaries<br />

(mm). This figure can be compared with Fig.17.2 (loo%*). In these experiments hMES = 139;<br />

hglutamate = 77.5; hacetate = 17.5; hamarant,, red = 0; hence Ah refers to the step height of the<br />

amaranth red zone.<br />

Fig.17.6. Isotachopherogram of a standard mixture of anions (Fig.12.5) without (A) and with (B) a<br />

counter flow of electrolyte, showing that several mixed zones disappear. The experiments were carried<br />

out in the operational system at pH 6 (Table 12.1). The differences in heights of the peaks (impurities)<br />

in the linear trace of the W detector should be noted. These zones are marked with asterisks. They<br />

are enriched during the time that the counter flow of electrolyte is applied (see section 10.5). A slight<br />

difference between the recordings from the W absorption detector and the conductivity detector is<br />

always obtained because the diameters of the probe and the PTFE narrow-bore tube are not the same.<br />

Moreover, differences between the traces of the conductivity and the UV absorption detectors can be<br />

expected in (A), because the W absorption detector is mounted closer to the injection point. The<br />

zone of carbonate, marked with a large asterisk, increases with time (see Chapter 9). A = Increasing UV<br />

absorption; R = increasing resistance; t = time.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!