210 PRACTICAL ASPECTS<br />

9.7. CONVERSION OF DATA MEASURED WITH DIFFERENT DETECTORS<br />

When different apparatus are used at different times, it is sometimes difficult to compare<br />

the results. For thermocouples and conductivity detectors, however, a simple method can<br />

be used to interconvert the results. We can utilize the situation that the signals obtained<br />

from one detector, when represented graphically as a function of the signals of another,<br />

give a nearly straight line. After constructing a calibration graph, all results can then be<br />

converted by means of the conversion graph. This graph is independent of the different<br />

operational systems as it simply represents a conversion of measured temperatures and<br />

conductivities, and so it can be used for all systems.<br />

-<br />

D<br />

Ir,<br />

150 200 250 3&0<br />

t.c.2<br />

Fig.9.11. Conversion graph for two different thermocouples (t.c.1 and t.c.2). * = Anionic species<br />

given in Table 9.4; O= cationic species given in Table 9.4.